Autoclave Tomy Sx700 Manual Best File
The best manual contains a step-by-step for biological validation using Geobacillus stearothermophilus spore strips. This is non-negotiable for ISO 17025 labs.
With the unit unpressurized and cold, manually pull the ring on the safety relief valve to ensure it moves freely and is not seized by mineral deposits. 5. Troubleshooting and Error Codes
Manually check the pressure relief valve to ensure it moves freely and is not clogged. autoclave tomy sx700 manual best
| Feature | Specification (SX-700U / SX-700) | | :--- | :--- | | | 69 L (SX-700U) / 79 L (SX-700) | | Chamber Dimensions (Ø x H) | 370 x 774 mm (SX-700) | | External Dimensions (W x D x H) | 410 x 528 x 1003 mm (SX-700) | | Operating Temperature Range | Sterilizing: 105 - 135°C / Heating: 45 - 104°C / Warming: 45 - 95°C | | Max Operating Pressure | 0.25 MPa | | Heat Source | 3 kW electric heater | | Cooling System | Standard dual-fan cooling system |
105°C to 135°C for sterilization; 45°C to 104°C for heating. Maximum Pressure: 0.25 MPa (approx. 36 psi). Power Requirements: AC 220-240V, 15A. The best manual contains a step-by-step for biological
The water level sensor detects insufficient water in the chamber. Turn off the unit, let it cool, and add distilled water up to the level plate.
: Prevents media from coagulating after the cycle. Maximum Pressure: 0
: Specifically designed to prevent "sudden boiling" by cooling samples gradually. Normal Sterilizing : For solid laboratory waste and equipment. Sterilizing & Warming